Thermocouple calibrator T254

Thermocouple calibrator T254

All Tinycal series instruments have a dual function: generation/simulation and measurement. The keys on the front panel of the instrument allow configuration for the generation of set values in engineering units or percentages. The panel keys also allow the generation in jumps that can be programmed from 1 to 30 seconds per jump, which is stored thanks to its 6 configuration keys. The simulation from the instrument – T254 and T264 – can be configured to simulate the sensor response (Pt100 or selected thermocouple type).

All instruments in the Tinycal family offer high precision and high resolution measurements which, together with their ease of use, high reliability and convenient transport, as well as their reduced weight and small size, make them basic standards in an instrumentation workshop.

738,00 €

738,00 €

Configuration may affect final price.

In Stock

Ready for dispatch

Worldwide Shipping

Expert Support Include

Features

  • Operating temperature: 0 to 50°C
  • In case of possible connection or manipulation errors, the measuring circuit is protected against short circuits and external voltages up to 30V.
  • Rechargeable batteries with long autonomy.

Advantages

  • Measurement/Generation with a single tool
  • High accuracy
  • High resolution

Applications

  • Thermocouple measurement and simulation
  • Calibration of thermocouple transmitters
  • Testing of temperature control loops

Technical specifications

Function Measurement and simulation
Generation/simulation Manual or automatic (stored steps)
Automatic step interval Programmable from 1 to 30 seconds
Temperature coefficient 0.003%/°C
Rechargeable batteries NiMH
Autonomy 14 hours
Charging time 16 hours
Battery charger 230 V, 50 Hz
Operating temperature 0 to 50 °C
Display LCD, 4½ digits, 10 mm high, blue LED backlight
Dimensions 83 × 152 × 33 mm
Weight 240 to 300 g
CE marking Yes
Warranty 2 years, extendable to 3 years
Signal Thermocouple
Thermocouple types B, E, J, K, N, R, S, T (IEC 584-1)
Measurement/simulation range −15.00 to 75.00 mV
Temperature ranges B: 100 to 1820 °C; E: -270 to 1000 °C; J: -210 to 1200 °C; K: -270 to 1372 °C; N: -270 to 1300 °C; R: -50 to 1768 °C; S: -50 to 1768 °C; T: -270 to 400 °C
Input impedance >10 MΩ
Accuracy at 23 °C ± 3 °C ±0.025% FS
Uncertainty at one year ±0.05% FS
Resolution 0.01 mV and 0.1 °C
Cold-junction compensation Internal or manual, 0 to 50 °C

PRODUCT CONFIGURATION

Preview the available configuration fields. Complete your selection and purchase on the DR72 product page.